Features
2, 4, 6 or 12 direct universal inputs: DC voltage, current and thermocouples |
High speed storage capability: up to 8 samples per second per channel. |
Multiple storage media options: 3 1/2" 1.44MB disk,or Compact Flash up to 2 GB. |
5.6" Brilliant TFT active matrix color display or 5" monochrome display. |
Intuitive Touchscreen Control with anti-glare coating. |
Splash resistant front panel |
Standard DIN enclosure: 144mm x 144mm bezel |
6 relay contact or solid state relay outputs/3 digital control input. |
Parallel printer output. |
Serial output options: RS232 or RS485. |
10BaseT Ethernet connection. |
Companion Software for unit configuration, communication, data analysis, print and export to standard spreadsheet/database applications. |
Efficiency, Flexibility and Data Integrity
Time is Money! Corporate down-sizing and cost cutting leaves you with less time to accomplish your goals. Data-Chart recorders are virtually maintenance free. No paper or pens to replace, no mechanical parts to wear out and because they are digital instruments, they requires less time to calibrate. This allows you to use your time more efficiently.
We’re Flexible. DataChart recorders are unparalleled in providing the highest level of flexibility of any paperless recorder made in the world. Universal inputs, networking capability, powerful math packages and a multitude of display choices allow you to display, record and communicate your data the way you want.
Because your data is so important Data Chart recorders were designed to be ultra-reliable. We chose the finest components available and combined them with a robust modular mechanical structure. Our optimized design means fewer components are needed. Fewer components means fewer failures.Up to 2 Megabytes of non-volatile memory keeps your data safe. You will never lose recorded data, even during a power outage. Data is downloaded automatically to your choice of removable media: 3.5” 1.44 Meg disk or CompactFlash™ card (up to 2 Gig in size).
Viewability, Setup and Display Modes
You can choose from more than 20 different display modes including trends, bargraphs and digital indicators. Select pen colors, background colors and much more. Our intuitive touchscreen control makes configuring the DC2000 a breeze. We make full use of our screen with a large, easy to follow menu system. The DC2000 has a brilliant 5.6” TFT active matrix color LCD display which is the largest of any 144mm square recorder. We’ve even added a special anti-glare coating to optimize viewability under any conditions.

|
Specifications
OPERATING
| Input Signals |
| DC Voltage: |
Linear, Industrial square root, logarithmic ± 150mV, ± 1.25V, ± 2.5V |
Accuracy: 0.06% |
| ± 12.5V and ± 25V |
Accuracy: 0.1% |
| DC Current: |
4-20mA, 0 to 20mA and 0 to 50 mA. |
Accuracy:0.15% using external 50 ohm 0.1% shunt |
| Dry Contact: |
Open = 0, Closed = 1 |
|
| External: |
Signals can be input via serial port (Modbus). |
|
| Thermocouple Resolution: 0.1°C, Cold Junction Reference: 0.5°C (0 to 50°C) Thermocouple burnout detection. |
| Thermocouple Info. |
DataChart® Specifications |
| Type |
Accuracy |
°C |
Accuracy |
°F |
Accuracy |
| J* |
0.1% |
J -101 to 1200°C |
± 1.5°C |
-150 to 2190°F |
± 3.0°F |
| K** |
0.1% |
K -101 to 1372°C |
± 1.5°C |
-150 to 2500°F |
± 3.0°F |
| T** |
0.2% |
T -101 to 400°C |
± 1.5°C |
-150 to 750°F |
± 3.0°F |
| E** |
0.11% |
E -101 to 1000°C |
± 1.5°C |
-150 to 1832°F |
± 3.0°F |
| R |
0.16% |
R -50 to 1768°C |
± 3.0°C |
-58 to 3200°F |
± 6.0°F |
| S |
0.17% |
S -50 to 1768°C |
± -3.0°C |
-58 to 3200°F |
± 6.0°F |
| B |
0.22% |
B 0 to 1820°C |
± 4.0°C |
32 to 3300°F |
± 7.0°F |
| C |
0.13% |
C 0 to 2400°C |
± 3.0°C |
32 to 4350°F |
± 6.0°F |
| N** |
0.10% |
N -101 to 1300°C |
± 1.5°C |
-150 to 2372°F |
± 3.0°F |
| * (± 2.5°C -210 to -100°C) **(± 2.5°C -270 to -100°C) |
| RTD's : Base accuracy 0.2% or 0.5°C (1°F). Resolution 0.1°C 2 or 3 wire connection. Cable compensation to ± 50 Ohm Open and short circuit detection. |
| 10 ohm Cu |
-70 to 170°C |
-94 to 338 °F |
| 100 ohm Pt 385 |
-220 to 850°C |
-364 to 1560 °F |
| 100 ohm Pt 392 |
-180 to 820°C |
-292 to 1500 °F |
| 200 ohm Pt 385 |
-220 to 400°C |
-364 to 750 °F |
| 200 ohm Pt 392 |
-180 to 400°C |
-292 to 750 °F |
| 120 ohm Ni |
-70 to 300°C |
-94 to 570 °F |
| 1000 ohm NI |
-60 to 209°C |
-76 to 408 °F |
| (DIN 43760) |
|
|
| Input Resolution |
0.0015% of full scale, 16 bit unless otherwise stated |
| Input Impedance |
> 10 Meg on 150mV, 1.25V and 2.5V ranges, >100 K on 5, 12.5, 25 Volt ranges. |
| Input Channels |
2, 4, 6 or 12 direct |
| Max Input |
50 Vdc |
| CMNR |
>100db, 50/60 Hz |
| Measurement Rate |
Measures all direct input channels every 125 milliseconds (each channel 8 times/second independent of no. of channels). |
| Math Functions |
+, -, x, /, logarithms, totalization, powers, averages, timers, and custom equations. |
| EMC Compliance |
Meets or exceeds the requirements of EMC 89/336/EEC |
|
RECORDING
| Recording Rates |
Selectable from 8/sec. to 10 minutes |
| Data Format |
Proprietary binary format for data security. |
| Data Storage |
Data stored in non-volatile RAM and recorded automatically, or on demand, to on board removable media. Full media format and verify capability. |
| |
Media |
Measurements |
Capacity |
| Removable |
3.5" Disk |
700,000 |
1.44MB |
| Compact Flash |
50 million |
Up to 2 GB |
| Internal |
1 Mb RAM (Non-Volatile) or 2 Mb RAM (Non-Volatile) |
| File Types |
Data files, Alarm and Event files, Configuration files, Language files. Multiple files of different names on a single disk. |
|
DISPLAY
Display Type Color: LED backlit Active Matrix TFT Liquid Crystal Display (5.6 inch) with touchscreen control.
Resolution 320 x 240 pixels.
Display Type Mono LED backlit STN Liquid Crystal Display (5.0 inch) with touchscreen control.
Resolution 240 x 128 pixels.
Display Modes Graphics (Trending vertical or horizontal), Bar Graphs (vertical or horizontal), Digital Meters, Alphanumeric Alarm and Event Data or combinations on a split screen. Review trended data. Search by time, date or signal value.
Virtual Chart Speed Programmable from 0.5in/hr to 600in/hr or 10mm.hr to 15,000mm/hr. Chart speed is independent of storage rate.
Display Windows Time/Date, Graphics (Bars, Large Digital, Trends), Disk Status, Systems Status, Menu Button Bar, Unit Indentification,Alarms/Events.
|
|
FEATURES
Touch Screen Touch sensitive screen with Button Bar for simple programming and easy operation.
Math Package Algebraic equations (basic math, powers, roots, natural and base 10 logarithms, exponentiation), peak monitoring, differentials, true rolling averages, time averages, gated timing, conditionals (Boolean logic), totalization, logarithmic scaling. 12 Programmable constants.
Buffer Internal 128k buffer (RAM) enables real time browsing of historic chart data independent of recorded data. Equivalent browse capability is approximately 560 pen inches. This represents about 5½ days for 4 pens at 1 inch/hour.
File Browse Any data file on disk can be browsed. File directory allows selection of different file names.
Disk Full Alarm User can set disk full threshold. Disk errors and alarms can be routed to contact closure.
Trend direction User selectable Horizontal or Vertical chart trending. "Pen" pointers for easy trace identification.
|
|
POWER
Power Requirements 100 to 240Vac, 50/60Hz or 125 to 300Vdc, 35VA max. Optional 24Vdc +/-15%.
Power Fail Protection Programmed parameters stored in non-volatile memory. Clock battery backed. Data retention time without power >12 months.Chart and alarm browse buffers stored in non-volatile memory.
Safety UL (3111-1) cUL (IEC1010-1) CE low voltage directive 73/23/EEC. Complies with EN 61010-1.
|
|
ENVIRONMENTAL
Temperature5°C to 40°C per UL3111-1/IEC1010-1 with disk drive. -10°C to 50°C with PCMCA drive or ZIPTM Drive.
Humidity 10% to 80% RH per UL3111-1/IEC1010-1.
Wash Down IP65 Front panel only.
|
|
OPTIONS
Alarm Contacts 3 or 6 isolated Form C, 3 amp @ 250Vac or 26 Vdc.
Solid State Relays 3 or 6, 0.5 amp @ 30Vdc.
Remote Inputs 3 isolated inputs, user selectable as dry contact or 5 to 12 Vdc (mech. relay), 12 to 24 Vdc (SS relay) activated.. Inputs share a common. Configurable for chart control, alarm acknowledge/reset, event markers, totalizer reset or logic input.
Communications ESD protected RS232 with full hand shaking. Supports modem or isolated RS485 port. Protocol: MODBUS RTU, MODBUS ASCII or serial printer port. Ethernet: 10BaseT. Unit may be remotely configured.
Printer Port Parallel printer port (25 pin D shell connector).
|
DIMENSIONS

click here to enlarge image
|
| |
|
Accessories
|
Part No.
|
Description
|
|
CFCR
|
CompactFlash Card Reader
|
|
SW-3T
|
Companion Software
|
|
Guardian
|
Guardian Software
|
|
MAS50R
|
50 ohm precision shunt resistor
|
| CC-8 |
Carry Case with Padded Nylon Strap |
| ALM-OUT6-2 |
6 Form C relay contact outputs (field installable) |
| ALM-OUT3-2 |
3 Form C relay contact outputs (field installable) |
| RS485/RS232 KIT |
RS485/232 Serial Line communications (field installable) |
|
Ethernet Kit
|
Ethernet communications board (field installable) |
| CAL-NIST |
N.I.S.T. Traceable Certificate with Data |
Memory Cards
|
Part No.
|
Description
|
|
MC256MBCF
|
256 Meg CompactFlash card
|
|
MC512MBCF
|
512 Meg CompactFlash card
|
|
MC1024MBCF
|
1 Gig CompactFlash card
|
|
MC2048MBCF
|
2 Gig CompactFlash card
|
Companion and Guardian Software Packages
Companion Software is a powerful and intuitive Windows based application that allows you to monitor real time data or review previously recorded data in graphic or tabular format, search files for specific events, link alarm and event files to trended data, print graphic or tabular files and export files to spreadsheet applications such as Excel.Recorder configurations are easily generated using Companion Software and can be downloaded to your recorders storage media and transferred to the recorder or transferred directly over ethernet or serial line communications. Monitor, configure and control up to 32 units with the RS485 Modbus option or control from remote locations using a modem connection. Whatever the application, Companion Software puts you in complete control.
Guardian Software allows you to create a complete single station data supervisory and storage system. If you need to monitor data in real time or if you require redundant data storage to a PC, Guardian Software is the solution.Multiple Data-Chart 2000’s can be placed on a standard ethernet or Modbus network along with your other plant instruments and monitored in real time. In addition, data can be stored on your local PC greatly improving data management and security!
 |
|
|