Support About Us Find a Distributor Warranty Policies & Condition of Sale Register Your Product Contact Us Literature ISO 9001/2000 Certified Home Monarch Instrument |
Media Storage Locations Rear Accessible: Internal:
Write Directly on the Screen
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| General
Input Resolution: 0.0015% of full scale, 16 bit unless otherwise stated
Input Impedance: >1 Mohm
Input Channels: 6 or 12 direct input plus 6 additional calculated channels
Maximum Input: 50Vdc
Isolation: Channel to Channel: 350Vdc or RMS AC
Channel to Chassis: 2000 Vdc or RMS AC
Isolation category II: Pollution Degree 2
Measurement Rate: 10 times per second on all direct input channels
Common Mode Noise Rejection: >100dB, 50/60 Hz, filter enabled
Normal Mode Noise Rejection: >50dB at 50/60 Hz, filter enabled
Math Functions: Fully programmable +, -, x, /, square root, sine, cosine, tangent, log,
totalization, powers, averages, conditional logic; AND, NOT, OR, +, >, <, gated timers. Can use live channels in calculation. Can define 6 constants and 6 functions per channel. Analog Inputs
DC Voltage: +/-125mV, +/-250mV, +/-500mV, +/-1.00V, +/-3.0V, +/-6.0V, +/-12.0V, +/-24.0V
Accuracy: Ranges to 1V +/-0.06%, Ranges > 1V +/-0.1%
DC Current: 4-20mA, 0-20mA, 10-50mA
Accuracy: +/-0.15% using external 50 ohm 0.1% 1/4 watt shunt
Thermocouple (Per ITS90)
Resolution: 0.1C, thermocouple burnout detection: Automatic
Reference junction compensation accuracy: +/-2.5 °C (0 °C to 50 °C)
Type Range (°C) Accuracy (°C) Range (°F) Accuracy (°F)
J -210 to -100°C +/-2.5 °C -340 to -150 °F +/-5 °F
-100 to 1200 °C +/- 1.5 °C -150 to 2190 °F +/- 3 °F
K -270 to -100 °C +/- 2.5 °C -450 to -150 °F +/- 5 °F
-100 to 1372 °C +/- 1.5 °C -150 to 250 °F +/- 3 °F
T -270 to -100 °C +/- 2.5 °C -450 to -150 °F +/- 5 °F
-100 to 400 °C +/- 1.5 °C -150 to 750 °F +/- 3 °F
E -270 to -100 °C +/- 2.5 °C -450 to -150 °F +/- 5 °F
-100 to 1000 °C +/- 1.5 °C -150 to 1832 °F +/- 3 °F
N -270 to -100 °C +/- 2.5 °C -450 to -150 °F +/- 5 °F
-100 to 1300 °C +/- 1.5 °C -150 to 2372 °F +/- 3 °F
S -50 to 1768 °C +/- 3 °C -58 to 3200 °F +/- 6 °F
B 0 to 1820 °C +/- 4 °C 32 to 3300 °F +/- 7 °F
RTD
Base Accuracy: 0.2% or 0.5 °C (1 °F). Resolution: 0.1 °C
2 or 3 wire connection. Cable compensation to +50 ohm. Open and short circuit detection.
Type Range °C Range °F
100 ohm Plt. 385 -220 to 85 °C -364 to 1560 °F
100 ohm Plt. 392 -180 to 820 °C -292 to 1500 °F
200 ohm Plt. 385 -220 to 400 °C -364 to 750 °F
200 ohm Plt. 392 -180 to 400 °C -292 to 750 °F
100 ohm Ni. -70 to 300 °C -94 to 570 °F
120 ohm Ni. -70 to 300 °C -94 to 570 °F
1000 ohm Ni. -60 to 209 °C -76 to 408 °F
10 ohm Cu. -70 to 170 °C -94 to 338 °F*
*0.5% +/- 0.5 °C
Frequency inputs (2 or 4 channels)
Range: 0 to 5,000Hz all channels, 0 to 10,000Hz 1 channel
Accuracy: 0.005% +/- 1 digit
Recording
Recording Rates: User programmable from 10 samples per second to 1 sample every 24 hours
Data Format: Proprietary encrypted format, User file naming
Data Storage Capacity: Data stored in non-volatile RAM and recorded automatically to:
Removable media types:
CompactFlash™ or USB drive to 4 GB
Internal media type:
SD card (secure digital) to 4 GB
File Types:Data files, alarm and event files, configuration files, language files, multiple files of different names on a single disk.
Display
Type: Color CCFL backlit Active Matrix TFT Liquid Crystal Display
Size: 5.6 inch diagonal, Resolution: 320 (W) x 240 (H) pixels
Interface: Resistive analog touch screen control
Display Builder: Allows user to create custom displays
Display Modes: Graphic trending (vertical or horizontal), Bar Graphs (vertical or horizontal), Digital Meter (large or small), Alphanumeric Alarm and Event log
Virtual Chart Speed: Programmable from 0.5 inch/hour to 600 inches/hour (10 mm/hr to 10,000 mm/hr)
Display Windows: Time/Date, Graphics (bars, large digital, trends) Disk Status, System Status, Menu Button Bar, Unit Indentification, Alarms/Events
Communications
Network: 10/100 Base T Ethernet per 802.3, RJ45 connection standard
Servers: Webserver supports http and ftp protocols, OPC server, Modbus over Ethernet server
Serial: Isolated RS485/RS232 Modbus Interface (option)
Power
Requirements: 100 to 240 Vac, 50/60Hz. 35 VA max. Optional 24 Vdc +/-15%
Power Fail Protection: Programmed parameters stored in non-volatile memory. Clock battery backed. Internal battery backup provides orderly shutdown and the ability to survive brownouts and short blackouts (<20 seconds)
Power Output: Optional isolated 24Vdc @ 120mA output
Input/Output
Digital I/O: 6 or 12 relay outputs, Form A (normally open SPST contacts) rated at 200 Vdc @ 0.5A Max, 2 digital control inputs +5 to +12Vdc @ 20mA (optional), Control inputs may be used for record start/stop, alarm acknowledge and channel reset functions
Safety and Environmental
Operating Range: 0 °C to 50 °C, 10% to 80% RH non-condensing
Protection: IP65 when mounted in panel
Safety: Meets the requirements of EN61010-1 when installed in accourdance with the intructions in the manual.
UL and cUL: Pending
EMC: Meets the requirements of EN61326:2003 and CE directive 89/336/EEC.
|
Options and Accessories
Memory Cards
Exhibitor SoftwareExhibitor Software is an extremely powerful set of tools that compliments the DC6000 and other OPC compliant devices. Incorporating functions to simplify data management via searching, reviewing, printing, or exporting historic data, Exhibitor Software allows real-time monitoring and recording independently as well, while historic recording is not affected. Also featured is the OPC Client which enables the user to build custom screens selecting various display elements and data from multiple OPC Servers, including devices other than DC6000’s. (Windows™ XP/2000 Compatible). Network Overview
|
ORDERING INFORMATIONPlease contact an authorized Monarch Instrument distributor in your area to place an order for the DataChart 6000 |
| Software |
| Data Sheets |
User Manuals |